Landauer’s NEUTRAK Dosimeter

IPLUS® and Neutrak®: High performance integrated neutron dosimeter

beta
gamma
neutrons
X
LANDAUER® provides you with IPLUS & Neutrak, a unique dosimeter incorporating the Neutrak detector in the IPLUS  badge. The dosimeter is used to measure X-rays, beta and gamma radiation, as well as thermal, intermediate and fast neutrons. Provide your team with a two-in-one dosimeter:
  • the lightest and most compact dosimeter on the market (17 g, 35 x 74 x 10 mm)
  • the only certified neutron dosimeter in France covering the energy range of thermal neutrons at 16.5 MeV
  • tested by the National Physical Laboratory (NPL), an accredited and independent laboratory in the UK
  • easy to wear and ready for use at the chest, wrist or ankle
  • compliant with EN 62387:2016 (IPLUS) and ISO 21909-1:2015 (Neutrak) standards

Technical Specifications

Technical Performance

IPLUS with Neutrak complies with the following standards:

  • EN 62387:2016 Radiation protection instrumentation – Passive Integrating Dosimetry Systems For Environmental And Personal Monitoring of photon and beta radiation
  • ISO 21909-1:2015 – Passive neutron dosimetry systems – Performance and test requirements for personal dosimetry
DosimeterIPLUSNeutrak-TNeutrak -J
ManufacturerLANDAUER (designed and manufactured in our laboratory in France)
Type of radiation measuredPhotons (X and gamma radiation) and betaNeutrons
Dose equivalentHp(10) and  Hp(0,07)Hp(10) chest and wrist
DetectorGN, new filter holder housingCR-39 ( PADC: poly-allyl diglycol carbonate)
TechnologyCarbon-doped aluminium oxide, Al2O3:CSSNTD (solid-state nuclear track detector) – Etched track detector
Minimum reporting value0.05 mSv0.1 mSv
ConverterN/APolyethylene and Teflon® doped with Boron 10Polyethylene
Linearity response0.05 mSv to 10 Sv – Standard deviation < 5 %0.1 to 10 mSv (thermal neutron)0.1 to 250 mSv (144 keV to 16.5 MeV)
0.1 to 250 mSv (from 144 keV to 16.5 MeV)
Energy response and related angular responsePhotons Hp(10) : ± 60° from 15 keV to 6 MeV
Photon : Hp(0,07) : ± 60° from 15 keV to 1.33 MeV
Beta: ± 45° from 0.24 to 0.8 MeV (Emean)
Thermal neutrons and from 144 keV to 16.5 MeV from 0° to ±60 °From 144 keV to 16.5 MeV from 0° to ±60 °td>

Dimensions: 35 x 74 x 10 mm
Weight: 17 g

Passive neutron dosimeter  integrated

Neutrak incorporated in the IPLUS dosimeter

Dimensions: 65 x 70 x 5 mm
Weight: 6,5 g

*Beware the bracelet must be ordered as well.

Neutrak & IPLUS adapted for wrist monitoring

  • COFRAC NF EN ISO/IEC 17025 Certification: 2005, scope n°1-1545, available on www.cofrac.fr
  • Approved in Belgium by the FANC (Federal Agency for Nuclear Control)
  • Participation in national and international intercomparisons
  • Dosimeter characterization completed by two standard independent laboratories: the Laboratoire National Henri Becquerel (LNHB) – CEA and the National Physical Laboratory (NPL) in the UK.